- Un-pack the mattress and place on top of the hospital bed and attach the adjustable straps around the four corners of the hospital bed frame.
- Make sure that the CPR release valve is at the left head side of the bed and is clear of the bed rails.
- Secure the pump to the footboard of the bed with the fold-away bed hanger brackets (LOCATED ON THE BACK OF THE PUMP.)
- Plug the pump into a grounded electrical outlet.
- Locate input panel on left side of the pump
- Line up notches on hose connector to the holes in the side panel.
- Twist connector to left until the lock slider is lined up with the slot in the panel. Slide the lock bar.
- FOR ATS-RC MODEL ONLY Connect hardwired remote to pump
- Plug the power cord into a grounded electrical outlet. The pump will turn on automatically.
- After the mattress is fully inflated, the caregiver can now transfer the patient onto the mattress.
- Push the AUTO FIRM button for faster inflation.
- The pump will change from AUTO FIRM to ALTERNATING mode automatically after 20 minutes.
- Static Mode: Push the Mode button to select static mode and adjust the Comfort Control by pressing the SOFT/FIRM button to achieve the maximum patient comfort. During this mode the system provides True Low Air Loss therapy. Perform a hand check by placing hand under the patient buttocks between air cells and bed frame.
- Alternate Mode: Push the Mode button to select Alternate mode and enable the 2-1 alternating pressure
- Alternation time can be adjusted by the CYCLE button. The time can be varied to 10, 15, 20 or 25 minutes. (If Static mode is selected the time window will not be illuminated)
- To fill mattress quickly use the optional orange inflator. Insert the inflator into the CPR Port located on the left head side of the bed. Twist the CPR port to the open position and fill until all bladders are firm. Once filled, twist the CPR port to the closed position. Allow system to run one complete 10 minute cycle.
- After 30 seconds the pump automatically engages a panel lock to ensure settings are not changed without authorization.
- To unlock panel, hold the button down until the orange light disappears.
Pump Setting Chart 35" & 42"
PATIENT WEIGHT LBS.
SETTING
90 - 125
1
120 - 185
2
155 - 225
3
180 - 265
4
225 - 300
5
275 - 350
6
300 - 375 (MAX 35")
7
350 - 425
8
425 - 525
9
525 - 600
10
Pump Setting Chart 48"
PATIENT WEIGHT LBS.
SETTING
125 - 175
1
150 - 225
2
225 - 265
3
250 - 350
4
300 - 375
5
350 - 425
6
425 - 500
7
475 - 550
8
550 - 625
9
625 - 750
10
Operating Procedures
- Once mattress is fully inflated, place patient on to the mattress. Allow the system to run a full cycle (10 min.).
- Adjust the comfort control knob to the patient’s desired level.
- A hand check can be performed by placing your hand under the patient on a deflated bladder. Make sure the patient is not touching the base of thebed. If this occurs increase the output by turning the control dial higher.
- In the event of cardiac arrest, twist the CPR, to release air. It will take less than 10 seconds to deflate.
- Once unit has cycled please refer to the weight chart. Keep in mind that this chart is just a guide. Patient comfort needs to be assessed as well.