TRADEWIND OPERATING INSTRUCTIONS - TRW-ATS & TRW-ATS-RC

Tradewind ATS-RC

 Tradewind ATS-RC with remote control shown above.

Setting up the System

  • - Un-pack the mattress and place on top of the hospital bed and attach the adjustable straps around the four corners of the hospital bed frame.
  • - Make sure that the CPR release valve is at the left head side of the bed and is clear of the bed rails.
  • - Secure the pump to the footboard of the bed with the fold-away bed hanger brackets (LOCATED ON THE BACK OF THE PUMP.)
  • - Plug the pump into a grounded electrical outlet.
  • - Locate input panel on left side of the pump
  • - Line up notches on hose connector to the holes in the side panel.
  • - Twist connector to left until the lock slider is lined up with the slot in the panel. Slide the lock bar.
  • - FOR ATS-RC MODEL ONLY Connect hardwired remote to pump
  • - Plug the power cord into a grounded electrical outlet. The pump will turn on automatically.
  • - After the mattress is fully inflated, the caregiver can now transfer the patient onto the mattress.
  • - Push the AUTO FIRM button for faster inflation.
  • - The pump will change from AUTO FIRM to ALTERNATING mode automatically after 20 minutes.
  • - Static Mode: Push the Mode button to select static mode and adjust the Comfort Control by pressing the SOFT/FIRM button to achieve the maximum  patient comfort. During this  mode the system provides True Low Air Loss therapy.  Perform a hand check by placing  hand under the patient buttocks between air cells and bed frame.
  • -  Alternate Mode: Push the Mode button to select Alternate mode and enable the 2-1 alternating pressure
  • - Alternation time can be adjusted by the CYCLE button.  The time can be varied to 10, 15,  20 or 25 minutes.  (If Static mode is selected the time window will not be illuminated) 
  • -  To fill mattress quickly use the optional orange inflator.  Insert the inflator into the CPR  Port located on the left head side of the bed.  Twist the CPR port to the open position and  fill until all bladders are firm.  Once filled, twist the CPR port to the closed position.   Allow system to run one complete 10 minute cycle.
  • -  After 30 seconds the pump automatically engages a panel lock to ensure settings are not  changed without authorization.
  • -  To unlock panel,  hold the button down until the orange light disappears.

 

 

Pump Setting Chart 35" & 42"

PATIENT WEIGHT LBS.

SETTING

90 - 125

1

120 - 185

2

155 - 225

3

180 - 265

4

225 - 300

5

275 - 350

6

300 - 375 (MAX 35")

7

350 - 425

8

425 - 525

9

525 - 600

10

 

 









Pump Setting Chart 48"

PATIENT WEIGHT LBS.

SETTING

125 - 175

1

150 - 225

2

225 - 265

3

250 - 350

4

300 - 375

5

350 - 425

6

425 - 500

7

475 - 550

8

550 - 625

9

625 - 750

10

 






Operating Procedures

  • - Once mattress is fully inflated, place patient on to the mattress. Allow the system to run a full cycle (10 min.).
  • - Adjust the comfort control knob to the patient’s desired level.
  • - A hand check can be performed by placing your hand under the patient on a deflated bladder. Make sure the patient is not touching the base of thebed. If this occurs increase the output by turning the control dial higher.
  • - In the event of cardiac arrest, twist the CPR, to release air. It will take less than 10 seconds to deflate.
  • - Once unit has cycled please refer to the weight chart. Keep in mind that this chart is just a guide. Patient comfort needs to be assessed as well.

 

Cleaning and Laundering Procedures

Disassembling the System

  • Unplug Pump from electrical outlet.
  • Disconnect mattress hoses from pump. 
  • Twist CPR Valve to open position to release air.
  • Insert orange deflator into the CPR Valve and evacuate air until mattress is totally flat.
  • Tuck hoses in mattress and roll it to fit into carry bag.

Safety Tips

  • Do not expose the pump or power cord to liquids of any kind
  • Always protect the system from open flames
  • Avoid the use of hypercarbonate or phenol based solutions in cleaning
  • Avoid the use of sharp objects near the mattress
  • Always store the mattress clean and dry
  • Consider the appropriate use of side rails on the patients bed